• DocumentCode
    1531920
  • Title

    Calibration of commercial power density meters at RF and microwave frequencies

  • Author

    Bostrom, R. ; Mild, K.H. ; Nilsson, G.

  • Author_Institution
    Nat. Primary Calibration Center, Arboga, Sweden
  • Issue
    2
  • fYear
    1986
  • fDate
    6/1/1986 12:00:00 AM
  • Firstpage
    111
  • Lastpage
    115
  • Abstract
    A calibration of 73 different probes from three different manufacturers has been carried out at 27 and 2450 MHz, respectively. The power density levels used were in the range 10-100 W/m2. The investigation was carried out at the Swedish National Primary Calibration Center, whose standards are traceable to the National Bureau of Standards, USA. The total uncertainty in the measurements is estimated to be ±0.7 dB for both 27 and 2450 MHz. Of the total number of instruments tested in this study, 5% did not work at all, although their owners thought they did. The results point to the necessity of a regular and relatively frequent calibration of the instruments to be used for monitoring of potentially hazardous electromagnetic fields.
  • Keywords
    calibration; field strength measurement; measurement standards; probes; safety; 10-100 W/m2; RF; calibration; field strength measurement; hazardous electromagnetic fields; microwave frequencies; monitoring; power density meters; probes; safety;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.1986.6499074
  • Filename
    6499074