DocumentCode :
1531920
Title :
Calibration of commercial power density meters at RF and microwave frequencies
Author :
Bostrom, R. ; Mild, K.H. ; Nilsson, G.
Author_Institution :
Nat. Primary Calibration Center, Arboga, Sweden
Issue :
2
fYear :
1986
fDate :
6/1/1986 12:00:00 AM
Firstpage :
111
Lastpage :
115
Abstract :
A calibration of 73 different probes from three different manufacturers has been carried out at 27 and 2450 MHz, respectively. The power density levels used were in the range 10-100 W/m2. The investigation was carried out at the Swedish National Primary Calibration Center, whose standards are traceable to the National Bureau of Standards, USA. The total uncertainty in the measurements is estimated to be ±0.7 dB for both 27 and 2450 MHz. Of the total number of instruments tested in this study, 5% did not work at all, although their owners thought they did. The results point to the necessity of a regular and relatively frequent calibration of the instruments to be used for monitoring of potentially hazardous electromagnetic fields.
Keywords :
calibration; field strength measurement; measurement standards; probes; safety; 10-100 W/m2; RF; calibration; field strength measurement; hazardous electromagnetic fields; microwave frequencies; monitoring; power density meters; probes; safety;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.1986.6499074
Filename :
6499074
Link To Document :
بازگشت