Improved Nearest Neighbor Methods for Gamma Photon Interaction Position Determination in Monolithic Scintillator PET Detectors
Author :
Van Dam, Herman T. ; Seifert, Stefan ; Vinke, Ruud ; Dendooven, Peter ; Löhner, Herbert ; Beekman, Freek J. ; Schaart, Dennis R.
Author_Institution :
Delft Univ. of Technol., Delft, Netherlands
Volume :
58
Issue :
5
fYear :
2011
Firstpage :
2139
Lastpage :
2147
Abstract :
Monolithic scintillator detectors have been shown to provide good performance and to have various practical advantages for use in PET systems. Excellent results for the gamma photon interaction position determination in these detectors have been obtained by means of the k -nearest neighbor (k -NN) method. However, the practical use of monolithic scintillator detectors and the k -NN method is hampered by the extensive calibration measurements and the long computation times. Therefore, several modified k-NN methods are investigated that facilitate as well as accelerate the calibration procedure, make the estimation algorithm more efficient, and reduce the number of reference events needed to obtain a given lateral (x,y) -resolution. These improved methods utilize the information contained in the calibration data more effectively. The alternative approaches were tested on a dataset measured with a SiPM-array-based monolithic LYSO detector. It appears that, depending on the number of reference events, ~ 10% to ~ 25% better spatial resolution can be obtained compared to the standard approach. Moreover, the methods amongst these that are equivalent to calibrating with a line source may allow for much faster and easier collection of the reference data. Finally, some of the improved methods yield essentially the same spatial resolution as the standard method using ~ 200 times less reference data, greatly reducing the time needed for both calibration and interaction position computation. Thus, using the improvements proposed in this work, the high spatial resolution obtainable with the k-NN method may come within practical reach and, furthermore, the calibration may no longer be a limiting factor for the application of monolithic scintillator detectors in PET scanners.
Keywords :
position sensitive particle detectors; positron emission tomography; solid scintillation detectors; PET scanners; PET systems; calibration data; gamma photon interaction; interaction position computation; k-NN method; monolithic LYSO detector; monolithic scintillator PET detectors; nearest neighbor methods; position determination; spatial resolution; standard method; Calibration; Crystals; Detectors; Nearest neighbor searches; Photonics; Radiation effects; Spatial resolution; Calibration; entry point; line source; monolithic scintillator detector; nearest neighbor method; position of interaction;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.2011.2150762
Filename :
5783323
Link To Document :
بازگشت