• DocumentCode
    1540961
  • Title

    Eddy current effects in MRI superconducting magnets

  • Author

    Badea, Eugene A. ; Craiu, Ovidiu

  • Author_Institution
    BSG Inc., Sugerland, TX, USA
  • Volume
    33
  • Issue
    2
  • fYear
    1997
  • fDate
    3/1/1997 12:00:00 AM
  • Firstpage
    1330
  • Lastpage
    1333
  • Abstract
    Magnetic resonance imaging (MRI), as well as some volume selected spectroscopy methods, use pulsed magnetic field gradients which induce multi-exponentially decaying eddy currents in all nonlaminated conductive parts of the superconducting magnets. This paper presents the analysis of the z gradient field distorsion due to the induced eddy currents and the corresponding correction in a 4 T/30 cm bore superferric self-shielded magnet
  • Keywords
    NMR imaging; eddy currents; superconducting magnets; 4 T; MRI superconducting magnets; NMR; eddy current effects; magnetic resonance imaging; multi-exponentially decaying eddy currents; nonlaminated conductive parts; pulsed magnetic field gradients; superferric self-shielded magnet; volume selected spectroscopy methods; z gradient field distorsion; Boring; Degradation; Eddy currents; Iron; Magnetic resonance imaging; Magnetic separation; Radio frequency; Solenoids; Superconducting coils; Superconducting magnets;
  • fLanguage
    English
  • Journal_Title
    Magnetics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9464
  • Type

    jour

  • DOI
    10.1109/20.582501
  • Filename
    582501