DocumentCode :
1542758
Title :
A simple method of energy calibration for thin plastic scintillator
Author :
Hohara, S. ; Saiho, F. ; Tanaka, J. ; Aoki, S. ; Uozumi, Y. ; Matoba, M.
Author_Institution :
Graduate Sch. of Eng., Kyushu Univ., Fukuoka, Japan
Volume :
48
Issue :
4
fYear :
2001
fDate :
8/1/2001 12:00:00 AM
Firstpage :
1172
Lastpage :
1176
Abstract :
A thin plastic scintillator is used as a trigger counter in a charged particle spectrometer. Since its Compton edge is not clear, it is difficult to test the thin plastic scintillator counter with a standard gamma-ray source. In this paper, a simple method is proposed to determine the Compton edge and the energy resolution of the thin plastic scintillator by fitting the simulated spectrum to the measured spectrum. The simulated spectrum was made with a Monte Carlo program utilizing the geometry of the plastic scintillator. This technique was examined with three different plastic scintillators
Keywords :
calibration; gamma-ray detection; solid scintillation detectors; Compton edge; Monte Carlo program; charged particle spectrometer; energy calibration; energy resolution; standard gamma-ray source; thin plastic scintillator; trigger counter; Calibration; Energy measurement; Energy resolution; Geometry; Monte Carlo methods; Plastics; Radiation detectors; Solid modeling; Spectroscopy; Testing;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.958745
Filename :
958745
Link To Document :
بازگشت