DocumentCode :
1545127
Title :
An automated SAR measurement system for compliance testing of personal wireless devices
Author :
Yu, Qishan ; Gandhi, Om P. ; Aronsson, Magnus ; Wu, Ding
Author_Institution :
Dept. of Electr. Eng., Utah Univ., Salt Lake City, UT, USA
Volume :
41
Issue :
3
fYear :
1999
fDate :
8/1/1999 12:00:00 AM
Firstpage :
234
Lastpage :
245
Abstract :
An automated specific absorption rate (SAR) measurement system has been developed for compliance testing of personal wireless devices. Unlike other systems, this system uses a model with a lossy ear-shaped protrusion and the accuracy of this experimental setup has been checked by comparing the peak 1-g SARs for ten cellular telephones, five each at 835 and 1900 MHz, with the results obtained using a 15-tissue anatomically based model with the finite-difference time-domain (FDTD) numerical electromagnetic technique. The SAR measurement system uses a three-dimensional (3-D) stepper motor to move a Narda Model 8021 E-field probe to measure the SAR distribution inside a head-shaped tissue-simulant phantom near the radiating device. The head and neck part of the model with an ear-shaped protrusion of 3 mm thickness is made of a lossy outer shell of 5-7 mm thickness of epoxy laced with KCl solution. The phantom is filled with appropriate frequency-specific fluids with measured electrical properties (dielectric constant and conductivity) that are close to the average for gray and white matters of the brain at the center frequencies of interest (835 and 1900 MHz). The implantable E-field probe is calibrated using the FDTD-calculated SAR variations for a slab model at two commonly used frequencies, 835 and 1900 MHz and is checked to have good isotropic characteristics (±0.23 dB) and a wide dynamic range (0.01-10 W/kg). The system is validated using a 223-mm-diameter sphere model. Peak 1-g SAR´s for ten telephones using different antennas are within ±1 dB of those obtained using the FDTD numerical method for the anatomical model of the head and neck region
Keywords :
biological effects of fields; brain; cellular radio; computerised instrumentation; electromagnetic fields; field strength measurement; finite difference time-domain analysis; safety; testing; 1900 MHz; 223 mm; 3 mm; 5 to 7 mm; 835 MHz; anatomical model; automated SAR measurement system; brain; cellular telephones; compliance testing; finite-difference time-domain analysis; implantable E-field probe; lossy ear-shaped protrusion; personal wireless devices; specific absorption rate; Automatic testing; Electromagnetic measurements; Electromagnetic modeling; Finite difference methods; Frequency; Imaging phantoms; Probes; System testing; Telephony; Time domain analysis;
fLanguage :
English
Journal_Title :
Electromagnetic Compatibility, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9375
Type :
jour
DOI :
10.1109/15.784158
Filename :
784158
Link To Document :
بازگشت