DocumentCode :
1546000
Title :
A closed loop ML algorithm for phase aberration correction in phased array imaging systems. II. Performance analysis [Ultrasound medical imaging]
Author :
Fortes, José Mauro P
Author_Institution :
Pontificia Univ. Catolica do Rio de Janeiro, Brazil
Volume :
44
Issue :
2
fYear :
1997
fDate :
3/1/1997 12:00:00 AM
Firstpage :
271
Lastpage :
286
Abstract :
For pt.I see ibid., vol.44, pp.259-70 (1997). The performance of the maximum likelihood closed loop circuits, proposed in part I for phase aberration correction in phased array imaging systems, is analysed. This analysis is helpful in designing and also in determining the tracking mode phase error variance performance of both closed loop circuits (1-D and 2-D). It is an approximate analysis, suitable under small errors conditions, and considered accurate for high values of the signal to noise ratio.
Keywords :
aberrations; acoustic signal processing; biomedical ultrasonics; closed loop systems; maximum likelihood estimation; medical signal processing; ultrasonic transducer arrays; 1-D; 2-D; closed loop ML algorithm; closed loop circuits; maximum likelihood closed loop circuits; performance analysis; phase aberration correction; phased array imaging systems; signal to noise ratio; small errors conditions; tracking mode phase error variance performance; ultrasound medical imaging; Analysis of variance; Circuits; Maximum likelihood estimation; Performance analysis; Phase estimation; Phase measurement; Phased arrays; Signal processing algorithms; Signal to noise ratio; Tracking loops;
fLanguage :
English
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher :
ieee
ISSN :
0885-3010
Type :
jour
DOI :
10.1109/58.585112
Filename :
585112
Link To Document :
بازگشت