• DocumentCode
    1546699
  • Title

    An examination of the spectra of the "Kick-Out" pulses for a proposed sampling oscilloscope calibration method

  • Author

    Paulter, Nicholas G. ; Larson, D.R.

  • Author_Institution
    Nat. Inst. of Stand. & Technol., Gaithersburg, MD, USA
  • Volume
    50
  • Issue
    5
  • fYear
    2001
  • fDate
    10/1/2001 12:00:00 AM
  • Firstpage
    1221
  • Lastpage
    1223
  • Abstract
    The spectra of the "kick-out" pulses used in a proposed sampler calibration method were examined. The proposed calibration method requires that the amplitude of these pulses be linear with an adjustment parameter called the offset voltage. The kick-out pulses were observed to have a nonlinear dependence on offset voltage, and this may affect the application of the proposed calibration method
  • Keywords
    calibration; digital storage oscilloscopes; pulse measurement; signal sampling; transient response; waveform analysis; Fourier transforms; difference waveform; impulse response; kick-out pulse spectra; kick-out waveform; linear amplitude; nonlinear dependence; nose-to-nose calibration method; offset voltage; pulse metrology; sampling oscilloscope; sampling-diode strobe pulse; Calibration; Diodes; Electric variables measurement; Metrology; Oscilloscopes; Photoconductivity; Pulse measurements; Sampling methods; Time domain analysis; Voltage;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.963187
  • Filename