DocumentCode
1546754
Title
Characterising differences between measurement and calibration wafer in probe-tip calibrations
Author
Carchon, G. ; Nauwelaers, B. ; De Raedt, W. ; Schreurs, Dominique ; Vandenberghe, Stefaan
Author_Institution
ESAT, Katholieke Univ., Leuven, Heverlee, Belgium
Volume
35
Issue
13
fYear
1999
fDate
6/24/1999 12:00:00 AM
Firstpage
1087
Lastpage
1088
Abstract
A general method to compensate for differences in probe-tip-to-line geometry and substrate permittivity between measurements and the calibration wafer is presented. The propagation constant, characteristic impedance and parameters of a general lumped-element equivalent discontinuity model are extracted at each frequency point based on measurements of two lines with different lengths.
Keywords
calibration; MMIC; calibration wafer; characteristic impedance; frequency point; lumped-element equivalent discontinuity model; probe-tip calibrations; probe-tip-to-line geometry; propagation constant; substrate permittivity;
fLanguage
English
Journal_Title
Electronics Letters
Publisher
iet
ISSN
0013-5194
Type
jour
DOI
10.1049/el:19990731
Filename
784544