Abstract :
A Monte Carlo study of the timing resolution of small BaF2 crystals for time-of-flight (TOF) positron emission tomography (PET) is presented. The interaction of 511-keV gammas, the emission and transport of ultraviolet light in the crystal, and the pulse formation in the photomultiplier are simulated. The individual contributions to the timing resolution and the overall effect are calculated and compared to experimental measurements. It is found that for small crystal lengths the choice of photomultiplier can have strong effect on the resolution
Keywords :
Monte Carlo methods; barium compounds; biomedical equipment; gamma-ray detection and measurement; radioisotope scanning and imaging; scintillation counters; time of flight spectra; BaF2; Monte Carlo study; TOF PET; gammas; photomultiplier; pulse formation; time of flight positron emission tomography; timing resolution; Biomedical engineering; Biomedical measurements; Detectors; Monte Carlo methods; Photomultipliers; Positron emission tomography; Signal resolution; Spatial resolution; Time measurement; Timing;