DocumentCode :
1554714
Title :
Dynamic SPECT imaging using a single camera rotation (dSPECT)
Author :
Farncombe, T. ; Celler, A. ; Noll, D. ; Maeght, J. ; Harrop, R.
Author_Institution :
Dept. of Phys., British Columbia Univ., Vancouver, BC, Canada
Volume :
46
Issue :
4
fYear :
1999
fDate :
8/1/1999 12:00:00 AM
Firstpage :
1055
Lastpage :
1061
Abstract :
Techniques to estimate dynamic parameters from the data acquired during a single rotation of a standard (single, dual and triple head) SPECT camera are being investigated. The performance of the constrained linear least squares (C-LS) method is being assessed. This method uses a simple assumption that the activity within each pixel at a given projection angle is less than or equal to the activity seen at the prior angle. In simulations, we have used an analytical heart model with single and bi-exponential activity decrease, with and without background activity and statistical noise in the data. Studies using a single, a 90° dual and a 120° triple head camera were modeled with a total acquisition time of 20 min, performing rotation over 1800. These simulations were followed by experimental scans of our dynamic phantom used with and without attenuation and subsequently a preliminary study involving a human subject was performed. Dynamic images from simulations and phantom experiments reconstruct well with an accuracy of reconstructed half-lives for most reconstructions within a 20% error range. Some streaking artifacts which are evident in images corresponding to early reconstruction times subside upon incorporating further constraints and regularizations
Keywords :
cardiology; image reconstruction; least squares approximations; medical image processing; single photon emission computed tomography; accuracy of reconstructed half-lives; analytical heart model; attenuation correction; background activity; bi-exponential activity decrease; constrained linear least squares method; dual head camera; dynamic SPECT imaging; dynamic phantom; image reconstruction; pixel activity; single camera rotation; single head camera; single-exponential activity decrease; statistical noise; streaking artifacts; triple head camera; Analytical models; Attenuation; Background noise; Cameras; Head; Heart; Image reconstruction; Imaging phantoms; Least squares methods; Parameter estimation;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.790826
Filename :
790826
Link To Document :