• DocumentCode
    1554717
  • Title

    Influence of malfunctioning block detectors on the calculation of single detector efficiencies in PET

  • Author

    Ferreira, N.C. ; Trébossen, R. ; Grégoire, M.C. ; Bendriem, B.

  • Author_Institution
    Service Hosp. Frederic Joliot, Orsay, France
  • Volume
    46
  • Issue
    4
  • fYear
    1999
  • fDate
    8/1/1999 12:00:00 AM
  • Firstpage
    1062
  • Lastpage
    1069
  • Abstract
    The authors studied the artifacts generated by the application of methods of crystal efficiency calculation based on the fan-sum, to situations of high variability of efficiency, namely the cases of a ring with defective blocks and acquisitions using narrow energy windows. They propose two efficiency calculation methods to reduce these artifacts: one method discriminates the crystals with an efficiency very different from the average, and then uses a modification of the fan-sum to estimate the efficiency; the other method calculates iteratively the fan-sum of each crystal, weighted by the efficiencies of the opposed detectors. The effect of using an off-center uniform cylinder is also discussed, since it influences the performance of the algorithms. The proposed algorithms are compared with the more conventional fan-sum method and with an algorithm used in a commercial tomograph, using measurements and simulations. The results show a significant reduction of the efficiency estimation errors, in severe situations where detectors in a ring have very different detection efficiency
  • Keywords
    biomedical equipment; gamma-ray detection; positron emission tomography; PET; algorithm performance; artifacts; commercial tomograph; efficiency estimation errors; fan-sum modification; malfunctioning block detectors influence; medical diagnostic imaging; nuclear medicine; off-center uniform cylinder; single detector efficiencies calculation; Detectors; Estimation error; Geometry; Iterative algorithms; Joining processes; Light scattering; Particle scattering; Photonic crystals; Positron emission tomography; Technological innovation;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.790827
  • Filename
    790827