• DocumentCode
    1554931
  • Title

    Constraints on tetrapolar tissue impedance measurements

  • Author

    Jones, D.M. ; Smallwood, R.H. ; Hose, D.R. ; Brown, B.H.

  • Author_Institution
    R. Hallamshire Hosp., Univ. of Sheffield, UK
  • Volume
    37
  • Issue
    25
  • fYear
    2001
  • fDate
    12/6/2001 12:00:00 AM
  • Firstpage
    1515
  • Lastpage
    1517
  • Abstract
    Impedance spectrum measurement using a tetrapolar surface probe provides a means of characterising biological tissue. However, parasitic capacitance and the presence of conductive fluid on the tissue surface impose an upper limit on both the frequency and impedance of measurements. The magnitude of these constraints has been evaluated by measurement and modelling
  • Keywords
    biological tissues; electric impedance imaging; biological tissue; conductive fluid; electrical impedance tomography; impedance spectrum measurement; parasitic capacitance; tetrapolar surface probe;
  • fLanguage
    English
  • Journal_Title
    Electronics Letters
  • Publisher
    iet
  • ISSN
    0013-5194
  • Type

    jour

  • DOI
    10.1049/el:20011034
  • Filename
    972185