Abstract :
We have designed and fabricated an anatomically accurate human head phantom that is capable of generating realistic electric scalp potential patterns. This phantom was developed for performance evaluation of new electroencephalography (EEG) caps, hardware, and measurement techniques that are designed for environments high in electromagnetic and mechanical noise. The phantom was fabricated using conductive composite materials that mimic the electrical and mechanical properties of scalp, skull, and brain. The phantom prototype was calibrated and testing was conducted using a 32-electrode EEG cap. Test results show that the phantom is able to generate diverse scalp potential patterns using a finite number of dipole antennas internal to the phantom. This phantom design could provide a valuable test platform for wearable EEG technology.
Keywords :
Dipole antennas; Electrodes; Electroencephalography; Humans; Phantoms; Scalp; Electroencephalography (EEG); electromagnetic interference (EMI); mechanical vibration; phantom; Brain; Electroencephalography; Electromagnetic Fields; Equipment Design; Head; Humans; Models, Biological; Phantoms, Imaging; Scalp; Skull;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.2012.2207434
Filename :
6239574
Link To Document :
بازگشت