DocumentCode :
1558219
Title :
A 32 x 32 capacitive micromachined ultrasonic transducer array manufactured in standard CMOS
Author :
Lemmerhirt, David F. ; Cheng, Xiaoyang ; White, Robert D. ; Rich, Collin A. ; Zhang, Man ; Fowlkes, J. Brian ; Kripfgans, Oliver D.
Author_Institution :
Sonetics Ultrasound Inc., Ann Arbor, MI, USA
Volume :
59
Issue :
7
fYear :
2012
fDate :
7/1/2012 12:00:00 AM
Firstpage :
1521
Lastpage :
1536
Abstract :
As ultrasound imagers become increasingly portable and lower cost, breakthroughs in transducer technology will be needed to provide high-resolution, real-time 3-D imaging while maintaining the affordability needed for portable systems. This paper presents a 32 × 32 ultrasound array prototype, manufactured using a CMUT-in-CMOS approach whereby ultrasonic transducer elements and readout circuits are integrated on a single chip using a standard integrated circuit manufacturing process in a commercial CMOS foundry. Only blanket wet-etch and sealing steps are added to complete the MEMS devices after the CMOS process. This process typically yields better than 99% working elements per array, with less than ±1.5 dB variation in receive sensitivity among the 1024 individually addressable elements. The CMUT pulseecho frequency response is typically centered at 2.1 MHz with a -6 dB fractional bandwidth of 60%, and elements are arranged on a 250 μm hexagonal grid (less than half-wavelength pitch). Multiplexers and CMOS buffers within the array are used to make on-chip routing manageable, reduce the number of physical output leads, and drive the transducer cable. The array has been interfaced to a commercial imager as well as a set of custom transmit and receive electronics, and volumetric images of nylon fishing line targets have been produced.
Keywords :
CMOS integrated circuits; buffer circuits; capacitive sensors; foundries; frequency response; image resolution; integrated circuit manufacture; micromechanical devices; readout electronics; real-time systems; ultrasonic transducer arrays; CMOS buffers; CMUT pulse echo; CMUT-in-CMOS approach; MEMS devices; blanket wet-etch step; capacitive micromachined ultrasonic transducer array; commercial CMOS foundry; frequency response; high-resolution 3D imaging; integrated circuit manufacturing; multiplexers; nylon fishing line targets; on-chip routing; readout circuits; real-time 3D imaging; sealing step; standard CMOS; ultrasound array prototype; Arrays; Electrodes; Finite element methods; Imaging; Metals; Transducers; Ultrasonic imaging; Computer-Aided Design; Electric Capacitance; Equipment Design; Equipment Failure Analysis; Image Enhancement; Image Interpretation, Computer-Assisted; Miniaturization; Phantoms, Imaging; Transducers; Transistors, Electronic; Ultrasonography;
fLanguage :
English
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher :
ieee
ISSN :
0885-3010
Type :
jour
DOI :
10.1109/TUFFC.2012.2352
Filename :
6242808
Link To Document :
بازگشت