Title :
Cross-plane scattering correction-point source deconvolution in PET
Author :
Shao, Lingxiong ; Karp, Joel S.
Author_Institution :
Dept. of Radiol., Pennsylvania Univ., Philadelphia, PA, USA
fDate :
9/1/1991 12:00:00 AM
Abstract :
The authors propose a new 2-D point source scattering deconvolution method. The cross-plane scattering is incorporated into the algorithm by modeling a scattering point source function. In the model, the scattering dependence on axial and transaxial directions is reflected in the exponential fitting parameters, and these parameters are directly estimated from a limited number of measured point response functions. The results comparing the standard in-plane line source deconvolution to the cross-plane point source deconvolution show that for a small source the former technique overestimates the scatter fraction in the plane of the source and underestimates the scatter fraction in adjacent planes. In addition, the authors also propose a simple approximation technique for deconvolution
Keywords :
computerised tomography; radioisotope scanning and imaging; algorithm; axial direction; cross-plane scattering correction; deconvolution; exponential fitting parameters; medical diagnostic imaging; nuclear medicine; point response functions; point source deconvolution; positron emission tomography; scatter fraction; transaxial direction; Aging; Deconvolution; Gamma ray detection; Gamma ray detectors; Gamma rays; Object detection; Optical imaging; Parameter estimation; Positron emission tomography; Scattering parameters;
Journal_Title :
Medical Imaging, IEEE Transactions on