Development of Detection Algorithm of Fatal Arrhythmia for a New Implantable Cardioverter Defibrillator
Author :
Yoshizawa, Makoto ; Inagaki, Masashi ; Uemura, Kazunori ; Sugimachi, Masaru ; Sunagawa, Kenji
Author_Institution :
Res. Div. on Adv. Information Tech., Tohoku Univ., Sendai
fYear :
2006
Firstpage :
3909
Lastpage :
3912
Abstract :
This study discusses the validity of new methods of detecting fatal arrhythmia used for ICDs. The method utilizes the steady state gain of the system from the ECG signal to the ventricular volume signal. After analyzing the data measured in a dog, it has been shown that the system can be approximated by a second order resonant system. Moreover, it has been indicated that the batch-type least squares method with the short data window shifting every time the R-wave occurs is better rather than a few versions of recursive algorithms for identifying time-varying parameters
Keywords :
defibrillators; electrocardiography; least squares approximations; medical signal detection; medical signal processing; prosthetics; ECG; R-wave; batch-type least squares method; detection algorithm; dog; fatal arrhythmia; implantable cardioverter defibrillator; second order resonant system; time-varying parameters; ventricular volume signal; Biomedical engineering; Cardiology; Detection algorithms; Electrocardiography; Heart beat; Laplace equations; Signal processing; Steady-state; Time varying systems; Transfer functions;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 2005. IEEE-EMBS 2005. 27th Annual International Conference of the
Conference_Location :
Shanghai
Print_ISBN :
0-7803-8741-4
Type :
conf
DOI :
10.1109/IEMBS.2005.1615316
Filename :
1615316
Link To Document :
بازگشت