• DocumentCode
    1570999
  • Title

    Development of Detection Algorithm of Fatal Arrhythmia for a New Implantable Cardioverter Defibrillator

  • Author

    Yoshizawa, Makoto ; Inagaki, Masashi ; Uemura, Kazunori ; Sugimachi, Masaru ; Sunagawa, Kenji

  • Author_Institution
    Res. Div. on Adv. Information Tech., Tohoku Univ., Sendai
  • fYear
    2006
  • Firstpage
    3909
  • Lastpage
    3912
  • Abstract
    This study discusses the validity of new methods of detecting fatal arrhythmia used for ICDs. The method utilizes the steady state gain of the system from the ECG signal to the ventricular volume signal. After analyzing the data measured in a dog, it has been shown that the system can be approximated by a second order resonant system. Moreover, it has been indicated that the batch-type least squares method with the short data window shifting every time the R-wave occurs is better rather than a few versions of recursive algorithms for identifying time-varying parameters
  • Keywords
    defibrillators; electrocardiography; least squares approximations; medical signal detection; medical signal processing; prosthetics; ECG; R-wave; batch-type least squares method; detection algorithm; dog; fatal arrhythmia; implantable cardioverter defibrillator; second order resonant system; time-varying parameters; ventricular volume signal; Biomedical engineering; Cardiology; Detection algorithms; Electrocardiography; Heart beat; Laplace equations; Signal processing; Steady-state; Time varying systems; Transfer functions;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2005. IEEE-EMBS 2005. 27th Annual International Conference of the
  • Conference_Location
    Shanghai
  • Print_ISBN
    0-7803-8741-4
  • Type

    conf

  • DOI
    10.1109/IEMBS.2005.1615316
  • Filename
    1615316