Design constraints for the inductive power and data link of an implanted body sensor
Author
Masuch, Jens ; Delgado-Restituto, Manuel
Author_Institution
Inst. of Microelectron. of Seville, Univ. of Seville, Seville, Spain
fYear
2009
Firstpage
425
Lastpage
428
Abstract
In this paper the basic system design constraints for the inductive power and data link are shown. The limiting factors for the choice of quality factors, coil inductance and carrier frequency are evaluated through analytical methods and simulations. The impact of the analysis on the required rectifier structure is also discussed. The analysis is focused on the development of a miniature biomedical sensor device whose outer dimensions shall not exceed 5-by-5 mm.
Keywords
Q-factor; biomedical electronics; biosensors; prosthetics; rectifiers; biomedical sensor device; carrier frequency; coil inductance; data link; implanted body sensor; inductive power; limiting factors; quality factors; rectifier structure; Biomedical computing; Biosensors; Coils; Inductance; Q factor; RLC circuits; Rectifiers; Resonance; Resonant frequency; Transponders;