DocumentCode
1577696
Title
A calorimeter for μW level optical power transfer standard
Author
Suzuki, Yasuyuki ; Chida, Naomichi ; Komiyama, Makoto
Author_Institution
Yokogawa Electr. Corp., Tokyo, Japan
fYear
1991
Firstpage
131
Lastpage
134
Abstract
A calorimeter than can measure μW-level optical power has been developed for an optical power transfer standard. It can measure the optical fiber simply by attaching an adapter. This calorimeter sensor uses a compensative absorber to reduce the influence of pressure fluctuation and temperature variation in the measurement room, and an ultra-low-noise preamplifier. With this calorimeter, the standard deviation of the measured value is in the range of 0.02% to 0.4% in 20-mW to 10-μW optical power measurements. An error evaluation for an optical power level of 10 μW yielded a two-sigma (two standard deviation) total uncertainty of 0.9%
Keywords
calorimeters; measurement standards; optical fibres; optical variables measurement; power measurement; calorimeter sensor; compensative absorber; error evaluation; measurement standards; optical fiber; optical power measurements; optical power transfer standard; pressure fluctuation; standard deviation; Fluctuations; Joining processes; Measurement standards; Optical fibers; Optical sensors; Power measurement; Preamplifiers; Pressure measurement; Standards development; Temperature sensors;
fLanguage
English
Publisher
ieee
Conference_Titel
Instrumentation and Measurement Technology Conference, 1991. IMTC-91. Conference Record., 8th IEEE
Conference_Location
Atlanta, GA
Print_ISBN
0-87942-579-2
Type
conf
DOI
10.1109/IMTC.1991.161558