• DocumentCode
    1579655
  • Title

    Implementation of a test system for advanced calibration and performance analyses of flickermeters

  • Author

    Gallo, Daniele ; Landi, Carmine ; Langella, Roberto ; Testa, Alfredo

  • Author_Institution
    Dipt. d Ingegneria dell´´Informazione, Universita di Napoli, Italy
  • Volume
    2
  • fYear
    2003
  • Firstpage
    1518
  • Abstract
    The paper presents an implementation of a test system for advanced calibration and performance verification of flickermeters. Test include not only the complete set of IEC standard test points, but also an array of different voltage fluctuations - related to common power quality issues - that may be encountered in actual use. This is done because it can be shown that different digital flickermeter implementations fully meeting the performance tests defined in IEC 61000-4-15 can still disagree significantly in some actual measurements. After the description and the performance analysis of the test system, a commercial instrument is tested and its response is presented.
  • Keywords
    calibration; computerised instrumentation; measurement standards; photodetectors; photometers; test equipment; IEC 61000-4-15; IEC standard test points; advanced calibration; commercial instrument testing; flickermeters; performance analysis; power quality issues; test system; voltage fluctuations; Calibration; Educational institutions; IEC standards; Instruments; Performance analysis; Power measurement; Power quality; Protocols; System testing; Voltage fluctuations;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 2003. IMTC '03. Proceedings of the 20th IEEE
  • ISSN
    1091-5281
  • Print_ISBN
    0-7803-7705-2
  • Type

    conf

  • DOI
    10.1109/IMTC.2003.1208003
  • Filename
    1208003