DocumentCode
1580109
Title
Fully Integrated and Low Power CMOS Amplifier for Neural Signal Recording
Author
Wang, Yufeng ; Wang, Zhigong ; Lü, Xiaoying ; Wang, Huiling
Author_Institution
Inst. of RF- & OE-ICs, Southeast Univ., Nanjing
fYear
2006
Firstpage
5250
Lastpage
5253
Abstract
This paper presents a fully integrated and low power CMOS amplifier for neural signal recording. It consists of a band-filtering preamp and a current-mode instrumentation amplifier. The design approach and the noise optimization are briefly discussed. A 0.6 mum CMOS technology is used for the design of the amplifier IC. The simulation shows that it has a pass band from 59 Hz to 12.8 kHz and a mid-band gain of 80 dB. The supply voltage is plusmn1.25V and the power consumption is 180 muW
Keywords
CMOS integrated circuits; biomedical equipment; microelectrodes; neurophysiology; noise; optimisation; preamplifiers; prosthetics; -1.25 V; 0.6 mum; 1.25 V; 180 muW; 59 Hz to 128 kHz; 80 dB; band-filtering preamp; current-mode instrumentation amplifier; fully integrated low power CMOS amplifier; neural prosthetics; neural signal recording; noise optimization; Biomedical electrodes; Biomedical engineering; CMOS technology; Capacitors; Circuit synthesis; Energy consumption; Instruments; Microelectrodes; Prosthetics; Signal design; CMOS; ENG; Neural signal recording; neural prosthesis;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2005. IEEE-EMBS 2005. 27th Annual International Conference of the
Conference_Location
Shanghai
Print_ISBN
0-7803-8741-4
Type
conf
DOI
10.1109/IEMBS.2005.1615663
Filename
1615663
Link To Document