Abstract :
A new cochlear implant system for animal experiments is proposed. The system uses a DSP-based sound processor to implement the six-channel continuous interleaved sampling (CIS) speech processing strategy. An infra-red telemetry is designed to remotely connect the power-demanding sound processor to receiver/stimulator attached on the animal. This enables the animal to move freely. The receiver/stimulator consumes about 40 mW, thus can operates for about 200 hours without changing battery
Keywords :
biomedical telemetry; digital signal processing chips; ear; medical signal processing; prosthetics; speech processing; DSP-based sound processor; behaving animals; cochlear implant system; infrared telemetry; power-demanding sound processor; receiver; six-channel continuous interleaved sampling; speech processing; stimulator; Animals; Band pass filters; Batteries; Circuits; Cochlear implants; Digital signal processing; Pulse width modulation; Signal processing; Telemetry; Transmitters;