• DocumentCode
    1580278
  • Title

    A room temperature set-up to compare the quantized Hall resistance with 1 Ω standards

  • Author

    Boella, G. ; Reedtz, G. Marullo

  • Author_Institution
    IEN Galileo Ferraris, Torino, Italy
  • fYear
    1991
  • Firstpage
    161
  • Lastpage
    164
  • Abstract
    A method has been developed to relate the mean value of the IEN primary group of 1-Ω standard resistors to the i=2 quantum level of the Hall resistance. It is based on an automated potentiometric system, a 36×360 Ω Hamon network, and a commercial Kusters current comparator bridge. A description of the equipment and of the measurement technique is given together with a detailed uncertainty budget. The typical fractional global uncertainty of the measurement of one standard deviation level is 5×10-8
  • Keywords
    electric resistance measurement; measurement standards; quantum Hall effect; 1 ohm; Hamon network; Kusters current comparator bridge; automated potentiometric system; fractional global uncertainty; measurement standards; quantized Hall resistance; room temperature; standard resistors; uncertainty budget; Bridges; Current measurement; Electrical resistance measurement; Electrons; Measurement standards; Performance evaluation; Resistors; Switches; Temperature; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 1991. IMTC-91. Conference Record., 8th IEEE
  • Conference_Location
    Atlanta, GA
  • Print_ISBN
    0-87942-579-2
  • Type

    conf

  • DOI
    10.1109/IMTC.1991.161567
  • Filename
    161567