DocumentCode :
1584671
Title :
Common Mode Noise Cancellation for Electrically Non-Contact ECG Measurement System on a Chair
Author :
Kim, Ko Keun ; Lim, Yong Kyu ; Park, Kwang Suk
Author_Institution :
Interdisciplinary Program in Med. & Biol. Eng., Seoul Nat. Univ.
fYear :
2006
Firstpage :
5881
Lastpage :
5883
Abstract :
Electrically non-contact ECG measurement system on a chair can be applied to a number of various fields for continuous health monitoring in daily life. However, the body is floated electrically for this system due to the capacitive electrodes and the floated body is very sensitive to the external noises or motion artifacts which affect the measurement system as the common mode noise. In this paper, the driven-seat-ground circuit similar to the driven-right-leg circuit is proposed to reduce the common mode noise. The analysis of this equivalent circuit is performed and the output signal waveforms are compared between with driven-seat-ground and with capacitive ground. As the results, the driven-seat-ground circuit improves significantly the properties of the fully capacitive ECG measurement system as the negative feedback
Keywords :
biomedical electrodes; electrocardiography; equivalent circuits; feedback; patient monitoring; capacitive electrodes; chair; common mode noise cancellation; continuous health monitoring; driven-right-leg circuit; driven-seat-ground circuit; electrically noncontact ECG measurement system; equivalent circuit; negative feedback; Circuit noise; Electric variables measurement; Electrocardiography; Electrodes; Monitoring; Motion measurement; Noise cancellation; Noise measurement; Noise reduction; Signal analysis;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 2005. IEEE-EMBS 2005. 27th Annual International Conference of the
Conference_Location :
Shanghai
Print_ISBN :
0-7803-8741-4
Type :
conf
DOI :
10.1109/IEMBS.2005.1615828
Filename :
1615828