DocumentCode
1586610
Title
New developments for the determination of the response function for a BC501A compact neutron spectrometer for fusion diagnostics
Author
Gagnon-Moisan, F. ; Reginatto, M. ; Zimbal, A.
Author_Institution
Phys.-Tech. Bundesanstalt, Braunschweig, Germany
fYear
2011
Firstpage
1
Lastpage
4
Abstract
The compact neutron spectrometer project was initiated to develop a device able to study fast neutrons produced by fusion reactions in a tokamak. The goal is to provide a fully characterized BC501A detector using a digital acquisition system developed by ENEA. The pulse height resolution and the response matrix of the detector are determined using experimental data acquired at the PTB facility. To achieve this goal, methods of data analysis involving entropy maximization and new software were developed. The methods are described and the current results for the pulse height resolution are presented.
Keywords
Tokamak devices; neutron spectrometers; plasma diagnostics; BC501A compact neutron spectrometer; BC501A detector; compact neutron spectrometer project; digital acquisition system; entropy maximization; fast neutrons; fusion diagnostics; fusion reactions; pulse height resolution; response function determination; tokamak; Cyclotrons; Detectors; Energy resolution; Entropy; Minimization; Neutrons; Software;
fLanguage
English
Publisher
ieee
Conference_Titel
Advancements in Nuclear Instrumentation Measurement Methods and their Applications (ANIMMA), 2011 2nd International Conference on
Conference_Location
Ghent
Print_ISBN
978-1-4577-0925-8
Type
conf
DOI
10.1109/ANIMMA.2011.6172866
Filename
6172866
Link To Document