DocumentCode
1586980
Title
Measurement of Elastic Properties of Blood Vessels
Author
Ilic, D. ; Moix, T. ; Sache, L. ; Bleuler, H. ; Lambercy, O. ; Ohta, M. ; Augsburger, L.
Author_Institution
Lab. de Syst. Robotiques, Ecole Polytech. Fed. de Lausanne
fYear
2006
Firstpage
6199
Lastpage
6202
Abstract
This paper is related to the measurements of the modulus of elasticity of an artery by studying the deformations due to the inflation of an angioplasty balloon catheter used for interventional radiology (IR) procedures. Various types of balloons are studied in order to characterize and compare their behaviors at the time of inflation. A test bench, consisting of an angioplasty balloon, a polyvinyl alcohol model and an actuator used to inflate a balloon, is developed for the realization of the experiments. The pressure-volume curve during the inflation of a balloon is observed. Elasticity modulus are derived with an analytical model of the measurement system. The results are then analyzed and compared to existing data from literature
Keywords
biomechanics; blood vessels; catheters; deformation; elastic moduli; surgery; actuator; angioplasty balloon catheter; artery; balloon inflation; blood vessels; deformations; elastic properties; elasticity modulus; interventional radiology; polyvinyl alcohol model; pressure-volume curve; Angioplasty; Arteries; Blood vessels; Catheters; Computational modeling; Elasticity; Radiology; Robots; Sensor systems; Testing;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2005. IEEE-EMBS 2005. 27th Annual International Conference of the
Conference_Location
Shanghai
Print_ISBN
0-7803-8741-4
Type
conf
DOI
10.1109/IEMBS.2005.1615911
Filename
1615911
Link To Document