  • Abstract
    This paper is related to the measurements of the modulus of elasticity of an artery by studying the deformations due to the inflation of an angioplasty balloon catheter used for interventional radiology (IR) procedures. Various types of balloons are studied in order to characterize and compare their behaviors at the time of inflation. A test bench, consisting of an angioplasty balloon, a polyvinyl alcohol model and an actuator used to inflate a balloon, is developed for the realization of the experiments. The pressure-volume curve during the inflation of a balloon is observed. Elasticity modulus are derived with an analytical model of the measurement system. The results are then analyzed and compared to existing data from literature
  • Keywords
    biomechanics; blood vessels; catheters; deformation; elastic moduli; surgery; actuator; angioplasty balloon catheter; artery; balloon inflation; blood vessels; deformations; elastic properties; elasticity modulus; interventional radiology; polyvinyl alcohol model; pressure-volume curve; Angioplasty; Arteries; Blood vessels; Catheters; Computational modeling; Elasticity; Radiology; Robots; Sensor systems; Testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2005. IEEE-EMBS 2005. 27th Annual International Conference of the
  • Conference_Location
    Shanghai
  • Print_ISBN
    0-7803-8741-4
  • Type

    conf

  • DOI
    10.1109/IEMBS.2005.1615911
  • Filename
    1615911