Title :
ADONIS, high count-rate HP-Ge γ spectrometry algorithm: Irradiated fuel assembly measurement
Author :
Pin, Patrick ; Barat, Eric ; Dautremer, Thomas ; Montagu, Thierry ; Normand, Stéphane
Author_Institution :
Nucl. Meas. Team, AREVA NC La Hague, Beaumont-Hague, France
Abstract :
ADONIS is a digital system for gamma-ray spectrometry, developed by CEA. This system achieves high count-rate gamma-ray spectrometry with correct dynamic dead-time correction, up to, at least, more than an incoming count rate of 3.106 events per second. An application of such a system at AREVA NC´s La Hague plant is the irradiated fuel scanning facility before reprocessing. The ADONIS system is presented, then the measurement set-up and, last, the measurement results with reference measurements.
Keywords :
gamma-ray spectrometers; germanium radiation detectors; ADONIS digital system; HP-Ge gamma spectrometry algorithm; dynamic dead-time correction; high count-rate gamma-ray spectrometry; high-purity germanium detector; irradiated fuel assembly measurement; irradiated fuel scanning facility; Apertures; Assembly; Detectors; Fuels; Instruments; Signal processing algorithms; Spectroscopy; gamma-ray spectrometry; high count-rate; high-purity germanium detector (HP-Ge); irradiated fuel measurement;
Conference_Titel :
Advancements in Nuclear Instrumentation Measurement Methods and their Applications (ANIMMA), 2011 2nd International Conference on
Conference_Location :
Ghent
Print_ISBN :
978-1-4577-0925-8
DOI :
10.1109/ANIMMA.2011.6172917