DocumentCode
1590082
Title
Relative Location Estimation of Power-Receiver from Transmitter for In Vivo Robotic Capsules
Author
Hwang, Ji-Su ; Kim, Jong-Dae ; Park, Chan-Young ; Ryu, Munho ; Kim, Jongwon
Author_Institution
Div. of Inf. & Commun. Eng., Hallym Univ., Chunchon
fYear
2006
Firstpage
6635
Lastpage
6637
Abstract
This paper presents an estimation method of the relative location of the power-receiver from the transmitter for in vivo robotic capsules. The proposed method utilizes the fact that the needed current for driving the power transmitter is increased when the receiver coil is inductively coupled with the transmitter coil. By investigating the current increase characteristic according to the relative locations of the receiver, we draw an estimation strategy on the relative location of power-receiver from the transmission coil
Keywords
endoscopes; medical robotics; receivers; transmitters; endoscopes; in vivo robotic capsules; power receiver; power transmitter; receiver coil; relative location estimation; transmitter coil; Coils; Couplings; Current measurement; Endoscopes; In vivo; Magnetic fields; Power amplifiers; Pulse width modulation; Robots; Transmitters;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2005. IEEE-EMBS 2005. 27th Annual International Conference of the
Conference_Location
Shanghai
Print_ISBN
0-7803-8741-4
Type
conf
DOI
10.1109/IEMBS.2005.1616023
Filename
1616023
Link To Document