Automated analysis of the effect of wall motion filters on quantitative color Doppler velocities
Author
Vandervoort, Pm ; Aghassi, DS ; Weyman, AE ; Thomas, JD
Author_Institution
Noninvasive Cardiac Lab., Boston, MA, USA
fYear
1992
Firstpage
639
Lastpage
642
Abstract
The authors have shown previously by in vitro flow simulations the effect of wall filtering on the accuracy of the quantitative color Doppler velocity display. To automate the analysis of wall filter impact on mean Doppler frequency estimation they developed an algorithm which allows evaluation of the effect of different wall filter frequency response characteristics, flow fields, detection thresholds, and Nyquist limits. This automated algorithm has proven to be very effective in explaining and understanding the importance of wall filter selection in the quantitative application of Doppler color flow mapping
Keywords
Doppler effect; biomedical ultrasonics; cardiology; haemodynamics; medical signal processing; ultrasonic velocity measurement; Nyquist limits; automated analysis; cardiology; detection thresholds; flow fields; intracardiac blood flow velocity; quantitative color Doppler velocities; vascular flow profiles; wall motion filters; Cardiology; Color; Cutoff frequency; Displays; Filtering; Filters; Frequency estimation; Frequency response; Hospitals; Motion analysis;