Title :
Study on a Fully Open Magnetic Resonance Imaging Device with a Slice Imaging Region
Author :
Baodong, Bai ; Yanli, Zhang ; Dexin, Xie ; Pingchou, Xia
Author_Institution :
Sch. of Electr. Eng., Shenyang Univ. of Tech.
Abstract :
A new unilateral permanent magnet configuration for magnetic resonance imaging (MRI) system producing a slice imaging region with field homogeneity is presented. It is a totally new problem about magnet design to build a slice imaging region with homogeneous magnetic field outside the magnet source. A novel design method with high precision is proposed here to probe the existence possibility of such a remote slice region and then is used to analyze the location and size of the slice region. The success of such a unilateral permanent magnet with a slice region for MRI builds a strong basis for some scientific research such as the construction of the unilateral MRI system with homogeneous magnetic field
Keywords :
biomedical MRI; biomedical equipment; permanent magnets; field homogeneity; fully open magnetic resonance imaging device; homogeneous magnetic field; slice imaging region; unilateral permanent magnet configuration; Coils; Electromagnets; Magnetic fields; Magnetic resonance imaging; Medical treatment; Minimally invasive surgery; Nuclear magnetic resonance; Permanent magnets; Probes; Superconducting magnets;
Conference_Titel :
Engineering in Medicine and Biology Society, 2005. IEEE-EMBS 2005. 27th Annual International Conference of the
Conference_Location :
Shanghai
Print_ISBN :
0-7803-8741-4
DOI :
10.1109/IEMBS.2005.1616119