• DocumentCode
    1592908
  • Title

    Alignment method to eliminate power line interference effect on signal averaging

  • Author

    Herrera-Bendezü, Luis G. ; Denys, B.G. ; Reddy, P.S.

  • Author_Institution
    Div. of Cardiology, Pittsburgh Univ., PA, USA
  • fYear
    1992
  • Firstpage
    447
  • Lastpage
    450
  • Abstract
    Signal averaging has been used to enhance signals with random noise. This technique relies on determining the fiducial points accurately. Power line noise affects fiducial point calculations significantly. The authors have developed an alignment method using cross correlation, template preselection, and signal processing that virtually eliminates fiducial point errors due to power line noise. Simulation studies are presented to illustrate the results. They used real electrocardiographic signals contaminated with 60 Hz and Gaussian noise
  • Keywords
    correlation methods; electrocardiography; interference suppression; medical signal processing; random noise; 60 Hz; ECG; Gaussian noise; alignment method; cross correlation; fiducial points; power line interference effect; random noise; real electrocardiographic signals; signal averaging; signal enhancement; signal processing; simulation studies; template preselection; Cardiology; Electrocardiography; Filtering; Gaussian noise; Interference elimination; Matched filters; Noise level; Noise reduction; Signal processing; Signal to noise ratio;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Computers in Cardiology 1992, Proceedings of
  • Conference_Location
    Durham, NC
  • Print_ISBN
    0-8186-3552-5
  • Type

    conf

  • DOI
    10.1109/CIC.1992.269498
  • Filename
    269498