DocumentCode :
1592969
Title :
Topology optimization in the passive shim for the permanent magnet NMR
Author :
Yang, Liu ; Yang, Minghao ; Dong, Lanlan
Author_Institution :
Coll. of Eng., China Agric. Univ., Beijing, China
fYear :
2010
Firstpage :
13
Lastpage :
16
Abstract :
An improved method for allowing a NMR magnet to meet its field homogeneity requirements has been developed. Using the adjoint variable method, topology design sensitivity is derived analytically. Coupled with continuum topology sensitivity analysis, steep descent method is employed to find an optimized distribution of the shimming block of iron for the permanent magnets NMR. Different free induction decay signal data shows the effectiveness of the proposed optimal design method.
Keywords :
NMR spectroscopy; optimisation; permanent magnets; sensitivity analysis; shims; topology; NMR magnet; adjoint variable method; optimal design method; permanent magnet; shimming block; steep descent method; topology design sensitivity; topology optimization; NMR; Permanent magnet; Sensitivity analysis; Topology optimization;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
World Automation Congress (WAC), 2010
Conference_Location :
Kobe
ISSN :
2154-4824
Print_ISBN :
978-1-4244-9673-0
Electronic_ISBN :
2154-4824
Type :
conf
Filename :
5665553
Link To Document :
بازگشت