    The occurrence of anodal stimulation during bipolar pacing in implantable pacemakers

  • Author

    Lu, Richard M.T. ; Steinhaus, Bruce M. ; Dawson, Albert K.

  • Author_Institution
    Telectronics Pacing Syst., Englewood, CO, USA
  • fYear
    1992
  • Firstpage
    495
  • Lastpage
    498
  • Abstract
    An attempt was made to determine the anodal threshold during bipolar pacing using different pacemaker leads from three manufacturers and to determine whether simultaneous anodal and cathodal stimulation were responsible for the observed morphological changes in the evoked electrogram. Seven animals were studied. The results indicated that the anodal threshold was not a definitive value, but was a probability curve. The studies determined that the lowest anodal threshold ranged from 1.8 V to >6.7 V. In addition, the occurrence of anodal stimulation was highly correlated with changes in the bipolar evoked electrogram. There was also a correlation between anodal stimulation threshold during bipolar pacing and the anodal unipolar pacing threshold
  • Keywords
    anodes; pacemakers; 1.8 to 6.7 V; animals; anodal stimulation; anodal threshold; anodal unipolar pacing threshold; bipolar pacing; cathodal stimulation; evoked electrogram; implantable pacemakers; morphological changes; pacemaker leads; probability curve; Animals; Cathodes; Delay; Dogs; Electrodes; Linear regression; Manufacturing; Pacemakers; Polarization; Robustness;