• DocumentCode
    1595817
  • Title

    Code Compression after Demodulation in Ultrasound Flow Imaging Systems

  • Author

    Zhao, Heng ; Liu, Kai ; Gao, Shangkai

  • Author_Institution
    Dept. of Biomed. Eng., Tsinghua Univ., Beijing
  • fYear
    2006
  • Firstpage
    7425
  • Lastpage
    7428
  • Abstract
    In this paper, an ultrasound flow imaging system with single-transmit coded excitation is studied. With the oversampled code sequence modulated by base code, code compression before or after demodulation is demonstrated to be equivalent. A coded excitation flow imaging system with relatively low analog-to-digital (AD) sampling rate is proposed. The performance of the system is evaluated by simulation results
  • Keywords
    biomedical equipment; biomedical ultrasonics; codes; demodulation; flow visualisation; haemodynamics; base code; code compression; coded excitation flow imaging system; demodulation; low analog-to-digital sampling rate; single-transmit coded excitation; ultrasound flow imaging systems; Biomedical imaging; Blood flow; Decoding; Demodulation; Frequency modulation; High-resolution imaging; Image resolution; Modulation coding; Radio frequency; Ultrasonic imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2005. IEEE-EMBS 2005. 27th Annual International Conference of the
  • Conference_Location
    Shanghai
  • Print_ISBN
    0-7803-8741-4
  • Type

    conf

  • DOI
    10.1109/IEMBS.2005.1616228
  • Filename
    1616228