• DocumentCode
    1600686
  • Title

    Development of high speed synchronous control system for real time 3D eye imaging equipment

  • Author

    KO, Jong-Sun ; Kim, Hyun-Sik

  • Author_Institution
    Dept. of Electr. Electron. & Comput. Eng., Dankook Univ., Seoul, South Korea
  • Volume
    1
  • fYear
    2004
  • Firstpage
    381
  • Abstract
    To show a retina shape and thickness on the computer monitor, a laser has been used in scanning laser ophthalmoscope (SLO) equipment using the traveling difference of the laser. This method requires exact synchronous control of laser traveling in optic system to show a clear 3-dimensional image of retina. To obtain this image, this exact synchronism is very important for making the perfect plane scanning. In this study, a synchronous control of the galvanometer to make 3-dimensional retina image is presented. For the more, a very simple mathematical model of the galvanometer is approved by experimental result.
  • Keywords
    biomedical optical imaging; computerised monitoring; control engineering computing; eye; galvanometers; laser applications in medicine; medical computing; computer monitor; galvanometer; high speed synchronous control system; optic system laser traveling; plane scanning; real time 3D eye imaging equipment; retina 3D image; scanning laser ophthalmoscope; Coils; Control systems; Laser modes; Optical control; Optical imaging; Real time systems; Reluctance motors; Retina; Sensor arrays; Synchronous motors;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Industrial Technology, 2004. IEEE ICIT '04. 2004 IEEE International Conference on
  • Print_ISBN
    0-7803-8662-0
  • Type

    conf

  • DOI
    10.1109/ICIT.2004.1490319
  • Filename
    1490319