DocumentCode
1604509
Title
Depth of anaesthesia patient models and control
Author
Abdulla, Shahab ; Wen, Peng
Author_Institution
Fac. of Eng. & Surveying, Univ. of Southern Queensland, Toowoomba, QLD, Australia
fYear
2011
Firstpage
37
Lastpage
41
Abstract
During surgery, the anaesthetist carefully controls the delivery of anaesthesia given to the patient in an effort to attain and maintain a consistent and adequate level of anaesthetic depth. The aim of this work is to design an internal model control (IMC) for anaesthesia depth, and to test it by simulation with clinical data. This study uses an internal model control structure for the adjustment of Bispectral Index (BIS). Performance of the two controllers has been studied for a step change in BIS, measured disturbances in the measured variables. In this study the simulation shows that the internal model control performed better than the PID controller.
Keywords
medical control systems; surgery; BIS; anaesthesia delivery control; anaesthesia patient models; anaesthetic depth; bispectral index; internal model control; surgery; Internal model control; depth of anaesthesia;
fLanguage
English
Publisher
ieee
Conference_Titel
Complex Medical Engineering (CME), 2011 IEEE/ICME International Conference on
Conference_Location
Harbin Heilongjiang
Print_ISBN
978-1-4244-9323-4
Type
conf
DOI
10.1109/ICCME.2011.5876701
Filename
5876701
Link To Document