DocumentCode :
1604537
Title :
A microfluidic system with fuzzy PID controller
Author :
Wei, Wei ; Guo, Shuxiang
Author_Institution :
Harbin Eng. Univ., Harbin, China
fYear :
2011
Firstpage :
47
Lastpage :
52
Abstract :
In this work, a novel micoropump actuated by EO (electroosmotic) actuators toward biomedical systems and drug delivery systems. The classical structure, which consists of an actuating unit, chambers and rectifiers, has been employed in the design. The EO actuators with the impulse input make the polydimethylsiloxane (PDMS) diaphragm which separates the two chambers reciprocate. So the working liquid can be pumped from inlet to outlet. Due to the large deflections of the PDMS diaphragm, the device is completely self-filling and able to pump liquid even operated in vertical situation. By using the indirect pumping mechanisms, the EO actuator can be used more widely without considering the properties of the working liquid. And a maximum flow rate of 133μl/min (using top water as working liquid) was obtained by an input voltage of 30V and an exciting frequency of 1.4Hz. In order to get a stable flow rate, a fuzzy PID control approach has been adopted in the controlling system. And some experiments were manipulated to show the characteristics of the PDMS diaphragm and the check valve. These experiments results demonstrate the diaphragm micropump performance and ability to deliver the fluid and drugs to biomedical systems.
Keywords :
bioMEMS; electrophoresis; fuzzy control; medical control systems; microactuators; microfluidics; micropumps; osmosis; three-term control; biomedical systems; drug delivery systems; electroosmotic actuators; fuzzy PID controller; micoropump; microfluidic system; Actuators; Micropumps; Diaphragm micropump; Drug delivery system; Electroosmotic actuaion; Fuzzy control;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Complex Medical Engineering (CME), 2011 IEEE/ICME International Conference on
Conference_Location :
Harbin Heilongjiang
Print_ISBN :
978-1-4244-9323-4
Type :
conf
DOI :
10.1109/ICCME.2011.5876703
Filename :
5876703
Link To Document :
بازگشت