• DocumentCode
    1608117
  • Title

    Improved Radio Frequency Receiver Front-end for Magnetic Resonance Imaging

  • Author

    Jia, Xueqin ; Wang, Xu ; Li, Qiang ; Li, Jinghong

  • Author_Institution
    BioMed. Eng., Northeastern Univ., Shenyang
  • fYear
    2006
  • Firstpage
    1328
  • Lastpage
    1331
  • Abstract
    Based on software radio theory, this paper focuses on researching a direct conversion structure for the radio frequency (RF) receiver front-end and tries to apply this flexible receiver front-end in MRI receiving system. In particular, we look at the conventional architecture of RF receiver front-ends of MRI; present architecture of a direct conversion for RF receiver front-end of MRI; outline the key aspects of designing such multi-channel and multi-mode front-end systems. In addition, based on this architecture, a practical RF front-end receiving system of MRI is given. Two parts are included in this system; one is the conventional receiver front-end the other is the direct conversion design for RF receiver front-end of MRI
  • Keywords
    biomedical MRI; radio receivers; software radio; direct conversion RF receiver front-end; magnetic resonance imaging; multi-channel front-end systems; multi-mode front-end systems; radio frequency receiver front-end; software radio theory; Band pass filters; Biomedical engineering; Computer architecture; Frequency conversion; Magnetic resonance imaging; Mixers; Radio frequency; Receivers; Signal processing; Software radio;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2005. IEEE-EMBS 2005. 27th Annual International Conference of the
  • Conference_Location
    Shanghai
  • Print_ISBN
    0-7803-8741-4
  • Type

    conf

  • DOI
    10.1109/IEMBS.2005.1616672
  • Filename
    1616672