DocumentCode
1608151
Title
A Method for Artifact Correction Due to Demodulation Phase Errors in Magnetic Resonance Imaging
Author
Liu, Zhengmin ; Wu, Haicheng ; Zhou, Heqin
Author_Institution
Dept. of Autom., Univ. of Sci. & Technol. of China, Hefei
fYear
2006
Firstpage
1336
Lastpage
1339
Abstract
In the process that MR signal is quadrature demodulated, the initial phases of sinusoid and cosinusoid used as demodulation reference signal exist errors, which results in artifacts in the image. It is typically overcome through precise control of the initial phases with the help of the special hardware, but the errors are not reduced completely. In this work, a method based on reference scan is proposed. Reference scan without frequency encoding gradient and phase encoding gradient is executed before FSE sequence, and an echo train is acquired. The phase errors of sinusoid and cosinusoid that are used to demodulate each echo in the reference echo train are calculated. Then the k-space data of image is corrected by these errors and artifacts are removed. The experiments using 0.35T MRI system demonstrate the effectiveness of this method
Keywords
biomedical MRI; medical image processing; 0.35 T; FSE sequence; MRI system; artifact correction; cosinusoid; demodulation phase errors; k-space data; magnetic resonance imaging; reference echo train; reference scan; sinusoid; Automation; Chemicals; Demodulation; Encoding; Error correction; Filters; Frequency; Hardware; Magnetic resonance imaging; Signal processing; Artifact correction; Magnetic resonance imaging; Phase error; Reference scan;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2005. IEEE-EMBS 2005. 27th Annual International Conference of the
Conference_Location
Shanghai
Print_ISBN
0-7803-8741-4
Type
conf
DOI
10.1109/IEMBS.2005.1616674
Filename
1616674
Link To Document