Abstract :
RF exposure guidelines recognize that an increase in local SAR due to metallic implants is possible, depending upon the size and shape of the implant and the position. This study describes the local SAR calculations for an anatomical human with metallic osteosynthesis plates embedded in the mandibular zone exposed to 2 GHz RF fields; FDTD analysis is applied. The main objective is to assess whether combination of two metallic plates aligned in parallel in the head region could be expected to cause SAR enhancements under near-field exposure condition at 2 GHz.
Keywords :
biomedical materials; bone; finite difference time-domain analysis; orthopaedics; phantoms; prosthetics; FDTD analysis; RF exposure guidelines; RF fields; SAR enhancements; anatomical phantom implantation; local SAR calculations; mandibular zone; metallic implants; metallic osteosynthesis plates; near-field exposure condition; Dipole antennas; Finite difference methods; Guidelines; Implants; Numerical models; Radio frequency; Time-domain analysis; finite-difference time-domain (FDTD) method; metallic implant; osteosynthesis-plate; radio-frequency (RF) exposure; specific absorption rate (SAR);
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Electromagnetics (iWEM), 2014 IEEE International Workshop on
Conference_Location :
Sapporo
Type :
conf
DOI :
10.1109/iWEM.2014.6963710
Filename :
6963710
Link To Document :
بازگشت