DocumentCode
1611012
Title
ECG Gated Ultrasonic Small Animal Imaging
Author
Liu, Jian-Hung ; Jeng, Geng-Shi ; Wu, Tung-Ke ; Li, Pai-Chi
Author_Institution
Dept. of Electr. Eng., Nat. Taiwan Univ., Taipei
fYear
2006
Firstpage
1762
Lastpage
1765
Abstract
Echocardiography is a routine clinical procedure to diagnose cardiac functions. The organic structure of the mouse is similar to that of human so that murine echocardiography has potentially become an effective tool for the assessment of human cardiovascular disease. However, clinical ultrasonic imaging systems are not suitable for murine cardiac imaging due to its limited spatial and temporal resolution. Thus, high frequency ultrasonic imaging (ges20 MHz) is necessary in order to provide spatial resolution at the order of 100 mum. Furthermore, due to the lack of transducer arrays at such a high frequency, single-element transducer with mechanical scanning is typically used. Thus the frame rate is insufficient for imaging the quick motion of the mouse. In this paper, a high frequency ultrasonic imaging system with electrocardiography gating is built in order to provide both high spatial resolution and high temporal effecting resolution. The system utilizes the R-wave trigger signal from murine electrocardiography. Image data are acquired in either the block scanning mode or the line scanning mode. In block scanning, murine cardiac images in systole and diastole can be retrospectively reconstructed with a short data acquisition time. In line scanning, on the other hand, images during the entire cardiac cycle can be obtained. It is demonstrated that the effective frame rate can be up to 2 kHz, which is only limited by the pulse repetition rate of the system
Keywords
cardiovascular system; echocardiography; electrocardiography; image reconstruction; image resolution; medical image processing; ECG gated ultrasonic small animal imaging; R-wave trigger signal; block scanning mode; cardiac cycle; diastole; electrocardiography gating; high frequency ultrasonic imaging system; high spatial resolution; high temporal effecting resolution; human cardiovascular disease; image reconstruction; line scanning mode; murine echocardiography; systole; Animals; Echocardiography; Electrocardiography; Frequency; High-resolution imaging; Humans; Image resolution; Mice; Spatial resolution; Ultrasonic imaging;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2005. IEEE-EMBS 2005. 27th Annual International Conference of the
Conference_Location
Shanghai
Print_ISBN
0-7803-8741-4
Type
conf
DOI
10.1109/IEMBS.2005.1616787
Filename
1616787
Link To Document