• DocumentCode
    161574
  • Title

    Wireless power transfer system applied to an active implantable medical device

  • Author

    Campi, Tommaso ; Cruciani, Silvano ; Feliziani, Mauro ; Hirata, Akihiko

  • Author_Institution
    Dept. of Ind. & Inf. Eng. & Econ., Univ. of L´Aquila, L´Aquila, Italy
  • fYear
    2014
  • fDate
    8-9 May 2014
  • Firstpage
    134
  • Lastpage
    137
  • Abstract
    This study deals with the design of a Wireless Power Transfer (WPT) system based on magnetic resonant coupling between two coils, whose secondary is located inside the human body and connected to a battery recharge system of an active implantable medical device (AIMD). A co-simulation tool is developed using circuit and electromagnetic field models to simulate the WPT system. Simulation results are obtained for different positions of the secondary coil and are compared with experiments.
  • Keywords
    coils; electromagnetic field theory; inductive power transmission; prosthetics; secondary cells; AIMD; WPT system; active implantable medical device; battery recharge system; circuit field models; cosimulation tool; electromagnetic field models; magnetic resonant coupling; secondary coil; wireless power transfer system; Biological tissues; Coils; Couplings; Integrated circuit modeling; Muscles; Skin; Wireless communication; Wireless power transfer (WPT); active implantable medical device (AIMD); battery recharging; magnetic field numerical solution;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Wireless Power Transfer Conference (WPTC), 2014 IEEE
  • Conference_Location
    Jeju
  • Type

    conf

  • DOI
    10.1109/WPT.2014.6839612
  • Filename
    6839612