DocumentCode :
1619728
Title :
Improvement of mass standard up to 20 kg by multiple calibration method
Author :
Ueki, M. ; Mizushima, S. ; Sun, Jain-Xin ; Ueda, K.
Author_Institution :
Nat. Metrol. Inst., AIST, Tsukuba, Japan
Volume :
1
fYear :
2004
Firstpage :
450
Abstract :
Aiming at improving the mass standards, calibrations of standard weights of 1 kg to 20 kg have been performed by using a multiple calibration method. The mass comparisons of standard weights are made by three mass comparators installed in an air-tight chamber. The two automatic mass comparators used for measuring the masses of 2 kg to 20 kg are renewed in their self-centering system of weighing pans to improve the standard deviations of weighing indications. In addition, the lower setting limit of air density in the air-tight chamber is extended down to 0.2 kg/m/sup 3/ from 0.7 kg/m/sup 3/ and the reliabilities of evaluating the mass and the volume of weights are improved. As the reference standards of the multiple calibration method, two 1-kg stainless-steel weights, whose mass and volume are traceable respectively to the prototype kilogram and the silicon density standard, are used.
Keywords :
calibration; mass measurement; measurement standards; weight control; air-tight chamber; mass comparators; mass standard; multiple calibration method; reference standards; self-centering system; stainless-steel weights; standard deviation; weighing indications;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
SICE 2004 Annual Conference
Conference_Location :
Sapporo
Print_ISBN :
4-907764-22-7
Type :
conf
Filename :
1491443
Link To Document :
بازگشت