Title :
A study of reconstruction accuracy for a cardiac SPECT system with multi-segmental collimation
Author :
Yu, D.-C. ; Chang, W. ; Pan, T.-S.
Author_Institution :
Rush-Presbyterian-St. Lukes Med. Center, Chicago, IL, USA
Abstract :
To improve the geometric efficiency of cardiac SPECT imaging, we have previously proposed to use a ring geometry and a multi-segmental collimation. The proposed collimation consists of multiple parallel collimators with most of the segments focused on a small central region, where the patient heart should be positioned. This scheme provides an significantly increased detection efficiency for the central region, but at the expense of reduced efficiency for the surrounding background. The authors have used computer simulations to evaluate the implication of this scheme on the accuracy of the reconstructed cardiac images. Two imaging situations were simulated, one with the heart well placed in the center, the other with the heart shifted outward and partially outside the central region; a neighboring high uptake liver was also simulated. The images were reconstructed with ML-EM and OS-FM methods using a complete attenuation map. The results indicate the deviation caused by truncation is not significant and is not strongly dependent on the activity of the liver when the heart is well positioned within the central region. The distribution of activity in the myocardium reconstructed with ML-EM or OS-FM is not sensitive to the noisy projections sampled from the background. When the heart is positioned improperly, the image reconstructed from the hybrid emission (a combination of high-count projections the central region and low-count background projections) can restore the activity for the myocardium with increased noise variances in the section outside the central region
Keywords :
cardiology; digital simulation; image reconstruction; medical image processing; single photon emission computed tomography; cardiac SPECT system; complete attenuation map; computer simulations; heart; high-count projections; image central region; low-count background projections; medical diagnostic imaging; multisegmental collimation; myocardium; neighboring high uptake liver; noise variances; noisy projections; nuclear medicine; reconstruction accuracy; Background noise; Collimators; Computational modeling; Computer simulation; Geometry; Heart; Image reconstruction; Image segmentation; Liver; Myocardium;
Conference_Titel :
Nuclear Science Symposium, 1996. Conference Record., 1996 IEEE
Conference_Location :
Anaheim, CA
Print_ISBN :
0-7803-3534-1
DOI :
10.1109/NSSMIC.1996.587938