Optimization of the [18F]FDG protocol for clinical cerebral studies in 3D PET
Author :
Trébossen, R. ; Bendriem, B. ; Dupont, S. ; Dupel-Pottier, C. ; Semah, F. ; Remy, P.
Author_Institution :
CEA, Orsay, France
Volume :
3
fYear :
1996
Firstpage :
1872
Abstract :
The aim of this work was to optimize the [18F]FDG cerebral protocol for 3D PET acquisitions on the EXACT HR+. We compared the noise equivalent counts (NEC) and the quantitation of 2D and 3D acquisitions both performed in 7 patients. The 2D and 3D scans were acquired sequentially during respectively 20 and 5 minutes after an injection of 160 to 255 MBq of [18F]FDG. A plasma input function was measured in three patients. The NEC analysis indicated that the optimal dose of [18F]FDG was close to 200 MBq for a 3D brain acquisition. Using this dose, a 3.5-fold increase in NEC was obtained in 3D compared to 2D allowing thus the acquisition time to be reduced to 6 minutes without any loss of NEC. First results in three subjects indicated slightly higher regional cerebral metabolism rate of the glucose (rCMRglc) in 3D than in 2D: the slope of the linear regression onto the 3D versus 2D rCMRglc values equaled 1.07±0.008, 1.04±0.006 and 1.07±0.005 (p<0.0001). The discrepancy between 2D and 3D values may be due in part to an overestimation of the scatter in 2D and to an undercorrection for the scatter component originating from out of FOV sources in 3D
Keywords :
brain; data acquisition; dosimetry; image reconstruction; medical image processing; positron emission tomography; radioactive tracers; 160 to 255 MBq; 2D scans; 3D PET; 3D PET acquisitions; 3D brain acquisition; 3D scans; 6 min; EXACT HR+; F; FOV sources; [18F]FDG protocol optimization; acquisition time; clinical cerebral studies; glucose; linear regression; noise equivalent counts; optimal dose; plasma input function; quantitation; rCMRglc; regional cerebral metabolism rate; scatter component; Biochemistry; Calibration; Image reconstruction; Linear regression; National electric code; Plasma measurements; Positron emission tomography; Protocols; Scattering; Sugar;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium, 1996. Conference Record., 1996 IEEE
Conference_Location :
Anaheim, CA
ISSN :
1082-3654
Print_ISBN :
0-7803-3534-1
Type :
conf
DOI :
10.1109/NSSMIC.1996.587995
Filename :
587995
Link To Document :
بازگشت