Title :
Rapid millimetre-wave sampler response characterization to well beyond 120GHz using an improved nose-to-nose method
Author_Institution :
Microwave Technol. Center, Agilent Technol., Santa Rosa, CA, USA
Abstract :
Simple enhancements to the nose-to-nose method of characterizing samplers are described. The technique is applied to a millimetre-wave sampler. These allow sampler magnitude and phase response to be measured to 150GHz and above. Excellent agreement is found with magnitude measurements made using conventional traceable power sensors.
Keywords :
millimetre wave measurement; signal sampling; 120 to 150 GHz; magnitude response; millimetre-wave sampler; nose-to-nose method; phase response; Convolution; Delay; Diodes; Frequency measurement; Instruments; Length measurement; Microwave technology; Microwave theory and techniques; Pulse measurements; Sampling methods;
Conference_Titel :
Microwave Symposium Digest, 2003 IEEE MTT-S International
Conference_Location :
Philadelphia, PA, USA
Print_ISBN :
0-7803-7695-1
DOI :
10.1109/MWSYM.2003.1210423