• DocumentCode
    1627582
  • Title

    Performance of PTB´s caesiurn fountain CSF 1

  • Author

    Weyers, S. ; Schröder, R. ; Bauch, A.

  • Author_Institution
    Phys. Tech. Bundesanstalt, Braunschweig, Germany
  • fYear
    2001
  • fDate
    6/23/1905 12:00:00 AM
  • Firstpage
    46
  • Lastpage
    52
  • Abstract
    CSF1, a caesium atomic fountain frequency standard, has been developed and operated since the end of 1999 at the PTB. In an ongoing process, its uncertainty uB has been evaluated and varies between 1·10-15 and 2·10-15, depending on the mode of operation and the duration of the evaluation. Local comparisons with PTB´s conventional primary clocks, CS1 and CS2, and with NIST-F1 using GPS carrier phase and TWSTFT, revealed agreement within the respective uncertainties uB and the comparison uncertainty. Five measurements of the TAI scale interval with CSF1 have been performed, each lasting 15 days. Use of CSF1 for the realisation of. a free atomic time scale TAF(PTB) is discussed
  • Keywords
    atomic clocks; caesium; frequency standards; laser cooling; measurement uncertainty; CSF1 frequency standard; Cs; GPS carrier phase; International Atomic Time; PTB standard; SI unit; TAI scale interval; TWSTFT; caesium atomic fountain; cold atom collisions; free atomic time scale; frequency comparisons; laser-cooling; local comparisons; microwave cavity; primary frequency standard; uncertainty; Atom lasers; Atomic beams; Atomic measurements; Clocks; Frequency; Laser beams; Performance evaluation; Stationary state; Titanium; Uncertainty;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Frequency Control Symposium and PDA Exhibition, 2001. Proceedings of the 2001 IEEE International
  • Conference_Location
    Seattle, WA
  • ISSN
    1075-6787
  • Print_ISBN
    0-7803-7028-7
  • Type

    conf

  • DOI
    10.1109/FREQ.2001.956158
  • Filename
    956158