Abstract
The paper describes the mathematical model of a coaxially mounted thermal voltage converter. Results of simulations are compared with results of experiments conducted in the 1 kHz-1000 kHz frequency range
Keywords
calibration; measurement standards; voltage measurement; 1 to 1000 kHz; LF AC-DC difference; calibration; coaxially mounted thermal voltage converter; mathematical model; simulations; transfer standards; Analog-digital conversion; Coaxial components; Conductors; Connectors; Councils; Frequency conversion; Inductance; Testing; Uncertainty; Voltage;