• DocumentCode
    1656276
  • Title

    A calibration method of the LFB acoustic microscope system using isotropic standard specimens

  • Author

    Kushibiki, J. ; Wakahara, Toru ; Kobayashi, T. ; Chubachi, N.

  • Author_Institution
    Dept. of Electr. Eng., Tohoku Univ., Sendai, Japan
  • fYear
    1992
  • Firstpage
    719
  • Abstract
    Absolute accuracy of the line-focus beam (LFB) acoustic microscope system is investigated for measurements of the leaky surface acoustic wave (LSAW) velocity and attenuation, and a method of system calibration is proposed. In order to discuss the accuracy, it is necessary to introduce a standard specimen whose bulk acoustic properties, viz., the independent elastic constants and density, are measured with high accuracy. Three synthetic silica glass substrates from different manufacturers are taken as isotropic standard specimens. The LSAW propagation characteristics, measured at a frequency of 225 MHz, are compared with the calculated results, using the measured bulk acoustic properties
  • Keywords
    acoustic microscopes; calibration; surface acoustic waves; ultrasonic velocity measurement; 225 MHz; SiO2; calibration method; isotropic standard specimens; leaky SAW attenuation; leaky SAW velocity; line-focus-beam acoustic microscope system; synthetic silica glass substrates; Acoustic beams; Acoustic measurements; Acoustic waves; Attenuation measurement; Calibration; Frequency measurement; Measurement standards; Microscopy; Surface acoustic waves; Velocity measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 1992. Proceedings., IEEE 1992
  • Conference_Location
    Tucson, AZ
  • Print_ISBN
    0-7803-0562-0
  • Type

    conf

  • DOI
    10.1109/ULTSYM.1992.275907
  • Filename
    275907