  • Abstract
    Heater-thermopile isolation of planar, thin film, multijunction thermal converter decreases with frequency. This causes an interaction with the measurement circuit, resulting in increased error of measurement of its AC-DC transfer difference. The paper discusses the sources of this error and methods of its reduction, illustrated by results of experiments.
  • Keywords
    comparators (circuits); convertors; measurement errors; thin film devices; transfer standards; voltage measurement; AC-DC transfer difference; AC-DC transfer standards; common mode voltage; heater-thermopile isolation; high-output-resistance standards; measurement error; multijunction thermal converter; nanovoltmeter; optimal comparator; planar thin film converters; primary standards; uncertainty; Analog-digital conversion; Circuit testing; Councils; Frequency conversion; Frequency measurement; Heat transfer; Instruments; Low voltage; UHF measurements; Uncertainty;