DocumentCode
1662908
Title
Samples above the median: a nonparametric method for intracardiac signal assessment
Author
Morris, M.M. ; Jenkins, J.M. ; DiCarlo, L.A.
Author_Institution
Dept. of Electr. Eng. & Comput. Sci., Michigan Univ., Ann Arbor, MI, USA
fYear
1996
Firstpage
69
Lastpage
72
Abstract
In this investigation, a nonparametric (distribution-free) detection scheme, Samples Above the Median (SAM), is devised as an arrhythmia discriminant for implantable cardioverter defibrillators which is highly sensitive and specific yet introduces no appreciable computational demand. No widely accepted probability density function (p.d.f.) for intracardiac electrogram signals has been determined. A nonparametric method does not utilize the p.d.f. SAM was tested on 30 patients each with documented episodes of normal sinus rhythm (NSR) and ventricular tachycardia (VT). SAM achieved an electrogram specificity of 0.992 (476/480) and an electrogram sensitivity of 0.981 (471/480). SAM achieved an arrhythmia sensitivity and specificity of 1 using a 12 of 16 (75%) electrograms per arrhythmia criterion.
Keywords
defibrillators; electrocardiography; medical signal processing; arrhythmia criterion; computational demand; distribution-free detection scheme; electrogram specificity; implantable cardioverter defibrillators; intracardiac electrogram signals; intracardiac signal assessment; nonparametric method; normal sinus rhythm; ventricular tachycardia; Batteries; Cardiac disease; Cardiology; Computational complexity; Distributed computing; Electric shock; Medical treatment; Morphology; Rhythm; Testing;
fLanguage
English
Publisher
ieee
Conference_Titel
Computers in Cardiology, 1996
Conference_Location
Indianapolis, IN, USA
ISSN
0276-6547
Print_ISBN
0-7803-3710-7
Type
conf
DOI
10.1109/CIC.1996.542475